Published in Medical Imaging 1996: Physiology and Function from Multidimensional Images, Eric A. Hoffman, Editor, Proc. SPIE 2709, 40-54 (1996).





Xenon Enhanced CT Imaging of Local Pulmonary Ventilation

Jehangir K. Tajik, Binh Q. Tran and Eric A. Hoffman

Division of Physiologic Imaging
Department of Radiology
University of Iowa College of Medicine
Iowa City, IA 52242


[ home | search | contact us ]




Table of Contents

Abstract
Introduction
Methods
Results
Discussion
Conclusion
Acknowledgements
References

1 Abstract


We are using the unique features of electron beam CT (EBCT) in conjunction with respiratory and cardiac gating to explore the use of non-radioactive xenon gas as a pulmonary ventilation contrast agent. The goal is to construct accurate and quantitative high-resolution maps of local pulmonary ventilation in humans. We are evaluating xenon-enhanced computed tomography in the pig model with dynamic tracer washout/dilution and single breath inhalation imaging protocols. Scanning is done via an EBCT scanner which offers 50 msec scan aperture speeds. CT attenuation coefficients (image gray scale value) show a linear increase with xenon concentration (r=0.99). We measure a 1.55 Hounsfield Unit (HU) enhancement (kV=130, mA=623) per percentage increase in xenon gas concentration giving an approximately 155 HU enhancement with 100% xenon gas concentration as measured in a plexiglass super-syringe. Early results indicate that a single breath (from functional residual capacity to total lung capacity) of 100% xenon gas provides an average 321.85 (SE) HU enhancement in the lung parenchyma (maximum 50 HU) and should not encounter unwanted xenon side effects. However, changes in lung density occuring during even short breath holds (as short as 10 seconds) may limit using a single breath technique to synchronous volumetric scanning, currently possible only with EBCT. Preliminary results indicate close agreement between measured regional xenon concentration-time curves and theoretical predictions for the same sample. More than 10 breaths with inspirations to as high as 25cmH2O airway pressure were needed to clear tracer from all lung regions and some regions had nearly linear rather than mono-exponential clearance curves. When regional parenchymal xenon concentration-time curves were analyzed, vertical gradients in ventilation and redistribution of ventilation at higher inspiratory flow rates were consistent with known pulmonary physiology. We present here a works in progress, showing results from two pigs illustrating the high resolution and detailed regional information obtainable with careful attention to cardiac and respiratory gating during a multi-breath washout protocol.

Keywords: xenon, pulmonary imaging, computed tomography, regional ventilation, physiologic imaging, electron beam CT, lung, pulmonary physiology


2 Introduction


Calculating regional pulmonary ventilation requires measuring the rate of delivery (wash-in) of a tracer to lung regions or the clearance rate (washout) of a tracer from lung regions. Early studies of pulmonary ventilation used radioactive isotopes of chemically inert gases such as 133xenon as the tracer and measured either the buildup of radioactivity (wash-in study) or the decrease in radioactivity (washout study) of lung zones over multiple breaths [2]. More recently, the enhanced spatial resolution provided by dynamic x-ray CT has been used clinically in conjunction with inhalation of stable (non-radioactive) 131xenon gas as a contrast agent to measure local pulmonary ventilation [14,1,15,16,20,8,5,6] and regional cerebral blood flow [11,19,7,13]. Though there have been several studies using dynamic x-ray CT in conjunction with stable xenon to calculate local pulmonary ventilation, there has been little if any information generated as to the accuracy and validity of these measurements or studies utilizing accurately controlled simultaneous respiratory and cardiac gating.

Stable (non-radioactive) xenon gas has a K-edge similar to iodine and should, therefore, be a potent x-ray attenuator and provide good contrast enhancement when used in conjunction with CT scanning[17]. Though xenon gas is soluble in blood and tissues[18], it is chemically inert and thus has minimal physiologic effects. Xenon gas has been demonstrated to have an anesthetic effect when concentrations of 40% or greater are breathed for prolonged periods[4].

We hypothesized that it may be possible to ascertain the relative distribution of pulmonary ventilation by having the subject take a single deep breath of a high xenon gas concentration mixture (i.e. 80% xenon/20% oxygen or 100% xenon). A single breath of a higher xenon gas mixture should not encounter unwanted xenon side effects as mean alveolar xenon concentration from a single breath of this gas mixture should not exceed 40%. Subtraction of identically gathered xenon enhanced and unenhanced images should yield information regarding the relative distribution of local pulmonary ventilation. The enhancement due to xenon normalized to regional alveolar gas volume, yields direct assessments of well ventilated vs. poorly ventilated lung regions. It is crucial to reproduce the same lung volume for the enhanced and unenhanced scans for proper image registration. Early xenon studies were poorly controlled for lung volume and thus suffered from serious mis-registration artifacts which, when uncorrected, leave the results uninterpretable for regional physiologic assessment.

Another method for quantitating regional pulmonary ventilation involves sampling regional tracer delivery or clearance curves followed by mathematical analysis of the concentration-time curves to calculate regional kinetic rate constants or turnover rates.

Previous work by others regarding x-ray CT assessments of local pulmonary ventilation have assumed the kinetics of xenon tracer gas filling or clearing the alveoli to follow the mono-exponential model described by Kety[12]. Kety's model of inert gas exchange at the lungs has been used because it describes the buildup or clearance of an inert gas as a simple exponential function whose slope, when plotted on semi-log scale, is directly proportional to the fractional ventilation (V/Vgas). Using this relationship, some previous CT studies of local pulmonary ventilation were limited to simply measuring two time points, before and after xenon inhalation, rather than accurately sampling the wash-in and/or wash-out curves.

where Ct is the concentration of inspired gas at time t, is the concentration (amount) of gas present in the lungs at time infinity (i.e., at equilibrium) and is related to volume, and k is the rate constant which relates alveolar ventilation to alveolar volume (V/Vgas).

The Kety model assumes that the inert tracer gas has zero solubility in blood. Though this assumption is obviously incorrect for xenon gas as demonstrated by its anesthetic effects, previous CT assessments have ignored xenon solubility. Moreover, the Kety model has several assumptions and gives an imperfect fit to experimental ventilation data from radioactive 133Xe studies, leading some investigators to develop complicated multi-compartment models of inert gas uptake and exchange to better fit the data. We believe that it may be possible to correct for xenon solubility in blood without resorting to complicated multi-compartment analysis by using the high temporal resolution capabilities unique to electron beam CT[3] (EBCT) to dynamically follow the xenon concentration in blood over time simultaneously with the alveolar xenon wash-in/wash-out study. Correction for xenon solubility in blood can be made by knowing the mixed venous xenon concentration as a function of time (i.e., xenon blood concentration entering the pulmonary circulation, sampled in the right ventricle) and the local pulmonary blood content (either directly from a corresponding perfusion scan or estimated by knowledge of the relationship between regional alveolar gas volume and regional blood volume if a perfusion scan is not done). Another possibility would be to assume that xenon is soluble in everything that is not air (i.e. pulmonary parenchyma tissue and blood) and this volume of air and non-air fraction can be directly computed from an unenhanced scan.

2.1 Ventilation Imaging Protocols

Our goal is to develop technique(s) using stable xenon gas as a contrast agent in conjunction with the unique capabilities of dynamic x-ray CT imaging that will allow constructing accurate, quantitative and valid high resolution maps of regional pulmonary ventilation in humans. One of the major drawbacks of previous x-ray CT assessments of local pulmonary ventilation has been the lack of accurate lung volume control over multiple breaths, so that misalignment between unenhanced and enhanced scans is considerable and a possible source of significant error. In contrast to previous work by others, we are developing the capability to sample the majority of the local alveolar xenon filling and/or clearance curves by accurately tracking and controlling lung volume in patients over multiple breaths during a stable xenon wash-in or washout study. Using the two modes of the EBCT scanner (high spatial resolution and high temporal resolution), a dynamic washout study is designed to image regional tracer gas clearance (time-intensity data) during tidal breathing and a single breath protocol gathers data for subtraction of a xenon enhanced scan from an identical scan prior to xenon inhalation. The multi-breath washout study provides information on the regional distribution of pulmonary ventilation under dynamic/tidal breathing conditions while the single breath protocol provides information on the distribution of ventilation under static conditions. The physiology of these two conditions is clearly dissimilar.

In order to generate accurate high-resolution maps of pulmonary ventilation from detailed regional measurements throughout the lungs, both imaging protocols employ respiratory and cardiac gating during image acquisition to reduce density changes due to cardiogenic motion artifact or respiratory movement rather than due to xenon enhancement. Both methods rely heavily on accurate registration of the multi-time point data and hence on accurate lung volume control to assure that scanning occurs at identical lung volumes.

In parallel with these studies we are developing the lung volume control equipment and software needed for human ventilation studies. In these initial animal studies, we were able to control airway pressure (hence lung volume) either via our computer controlled ventilator or by using a calibrated super-syringe during scanning.


3 Methods


3.1 Electron Beam Computed Tomography (EBCT)

Scanning protocols utilize an Imatron C-150L EBCT scanner[3] (figure 1). This scanner differs from conventional x-ray CT scanners in that a focused electron beam is swept along tungsten targets surrounding the subject instead of mechanically rotating an x-ray source around the subject. This setup allows for fast tomographic image acquisition with scan apertures on the order of 50-100 milliseconds as opposed to approximately 0.6-1.0 second scan apertures from most other commercial x-ray CT scanners. The fast scan times available with EBCT allow stop action images of the heart to be obtained and allow us to minimize lung density changes induced by the effect of cardiac motion.

 

 


Figure 1: Cross-sectional view of the Imatron EBCT scanner. See text for details. (With permission of Imatron, Inc.)

The EBCT scanner operates in a high temporal resolution mode (multi-time point imaging) or in a high spatial resolution mode (volumetric imaging). In the high temporal resolution mode, the electron beam is magnetically steered sequentially along 4 tungsten target rings encompassing 210 degrees about the subject. X-rays produced from each target ring are focused onto two detector rings so that up to 8 spatial levels may be acquired at each time point. Each pair of images is gathered within 54 milliseconds followed by a 8 millisecond pause as the beam is reset for the next target ring. All 8 spatial levels are therefore acquired within 224 milliseconds. The axial resolution for the multi-time point imaging mode is 1.5-3.0 mm if only a single tomographic slice is to be acquired over multiple time points or 7 mm if the multi-slice mode (up to 8 spatial tomographic levels covering 7cm) is used in conjunction with multi-time point imaging. This mode of scanning was used to acquire regional tracer clearance data (i.e. concentration-time data) during the multi-breath washout studies (see below).

In the high spatial resolution mode, a single tomographic slice (1.5-3.0 mm thickness) is acquired after which the patient table is advanced to the next tomographic image level. The higher spatial resolution of this mode is achieved by focusing the generated x-rays on a double-populated detector ring during a double sweep of a single target ring. The scan aperture in this mode is 100 milliseconds and images can be reconstructed using a high spatial frequency algorithm to enhance anatomic detail. This mode of scanning was used during the single breath imaging protocols (see below).

3.2 Non-Radioactive Xenon Gas as a Pulmonary Ventilation Contrast Agent

To ascertain the relationship between xenon concentration and x-ray attenuation (CT number or Hounsfield Unit), we imaged xenon concentrations of 100%, 80%, 60%, 30%, 15% and 0% (room air) in a plexiglass super-syringe with EBCT scanner settings of kV=130 and mA=623 during imaging. Mean HU was plotted against xenon gas concentration and the slope of the relationship determined by linear regression. This relationship is used to predict regional xenon enhancement after equilibration to a given percent xenon concentration as a verification of our initial measurements as is discussed below.

3.3 Animal Preparation

We report preliminary results here from two pigs. All animal studies were performed within guidelines for animal care adhered to by the American Physiological Society and the American Heart Association. The animal use protocols were approved in advance by the University of Iowa Institutional Animal Care and Use Committee.

Pigs were premedicated with I.M. ketamine (1 mg/kg) and acepromazine and anesthesia induced with bolus (2.5 mg/kg) propofol (Diprivan) in an ear vein. A cuffed endotracheal tube was placed and the pig mechanically ventilated (10 ml/kg) with a respirator. Under fluoroscopic guidance, pressure catheters (Millar) were placed in the right ventricle (pressure monitoring), right carotid artery (systemic blood pressure, blood gas monitoring) and right ventricular outflow tract (pulmonary artery pressure, non-ionic contrast injection, wedge pressure) for measurement and recording of physiologic information. After catheterization, pigs were transported to the EBCT facility and placed supine on the scanner table and connected to a CWE 9000 computer controlled respirator. ECG, airway pressure, systemic and right ventricular pressures were monitored using an IBM compatible computer equipped with a standard laboratory A/D converter board and recording software. The scanner x-ray on pulse was also recorded to allow exact correlation of physiologic and scanner events. Throughout the catheterization procedures and scanning protocols, deep anesthesia was maintained with propofol (12 mg/kg/hr) titrated to heart rate and reflexes. At the conclusion of scanning, the pig was euthanized with sodium pentobarbital (30 mg/kg) and concentrated KCl.

3.4 Imaging Protocols

Immediately preceeding each scanning protocol, the lungs were inflated to 25 cmH2O airway pressure 3-5 times to assure a constant volume history.

3.4.1 Single Breath Inhalation

The imaging protocol for the single breath study is illustrated in figure 2. The EBCT scanner is used in the high-spatial resolution mode (see above) so that a volumetric image spanning the apex to base extent of the lungs is gathered by advancing the patient table 3mm after the acquisition of each tomographic slice. Prior to the acquisition of the first slice, the subject inhales a single, deep breath of ambient air (unenhanced or baseline scan) after which respiration is suspended at a pre-selected lung inflation level (i.e. ambient, 15cmH2O, 25cmH2O airway pressure). After the onset of apnea, image acquisition begins with each tomographic level (slice) gathered with a 100 millisecond scan aperture ECG gated to the peak of the QRS complex. In the animal studies reported here, respiration remained suspended during the image acquisition period (approximately 40 seconds) and resumed immediately following the acquisition of the last image. The identical scan protocol is repeated substituting a single deep breath of a high xenon gas mixture (xenon enhanced scan) for the ambient air breath prior to imaging. In both instances, a stack of 40 3mm contiguous tomographic slices were gathered to generate a volumetric image data set of the lungs.

 

 


Figure 2: Sample imaging protocol for the single breath inhalation method. Top trace: scanner on pulses; Middle trace: ECG; Bottom trace: airway pressure along with scanner on signal (square wave). Note that scanning is triggered to the ECG signal during a period of extended apnea.

3.4.2 Multi-Breath Washout (Dynamic Protocol):

We sampled the washout curve at 10 or 20 time points with 1, 4 or 8 spatial levels acquired at each time point. It is of interest to determine the minimum and optimum number of time points (scans) needed to accurately represent the xenon clearance curve so that x-ray exposure/dosage can be minimized in human studies. It is critical that each time point (scan) be acquired at the same lung volume (i.e. slice registration) so that density changes within a parenchymal sample are due to local changes in xenon gas concentration rather than contaminated with density changes due to cardiogenic or ventilatory motion artifact.

For this multi-breath/dynamic protocol, both respiratory and cardiac gating is a requirement for proper image alignment at all time points. An example imaging protocol is illustrated in figure 3. For these studies, the EBCT scanner is used in the high temporal resolution mode, and the first image is acquired at equilibration of the lungs with a known xenon gas mixture. Cardiac gated image acquisition is accomplished by programming the scanner to acquire a multi-level data set every N+3 heart beats (where N is the number of beats required to accomplish a tidal breath) with a scan aperture of 50 milliseconds gated to 40% of the R-R interval. Concomitant respiratory gating was accomplished in the animal studies reported here via our customized computer controlled ventilator programmed to take a single breath during the N R-R intervals between successive image acquisitions. The timing (N+3) of the scanning was adjusted so that the animal had always returned to end expiration before the next set of images were gathered. This timing required customization for the animals passive expiration and for the tidal volume used in a given protocol.

 

 


Figure 3: Sample imaging protocol for the multi-breath washout method. Top trace: scanner on pulses; Middle trace: ECG; Bottom trace: airway pressure along with scanner on signal (square wave). Note that scanning is triggered to both the respiratory signal and to the QRS complex of the ECG.

3.5 Experimental Approach

Questions being addressed include cardiac mixing effects and temporal stability of the single breath tracer distribution, repeatability of the regional washout measurements and matching of the measured ventilation distributions with known pulmonary physiology.

3.5.1 Cardiac Mixing Effects and Temporal Stability of the Single Breath Tracer Distribution

Since the ECG gated volumetric data set in the single breath imaging protocol is gathered sequentially over roughly 40 seconds (largely determined by the heart rate and the scanners ability to follow a given heart rate), tracer xenon gas could be redistributed do to cardiogenic mixing leading to errors in the observed tracer distribution (i.e. tracer now redistributed due to mixing action of the heart on the lung parenchyma rather than representing the static distribution of ventilation.) To explore the possible effects of this phenomenon on the single breath change in HU distribution, we repeated the single breath imaging protocol using the high temporal resolution mode of EBCT scanning to follow possible density changes in a single tomographic section over time during breath hold at FRC after a breath of 100% xenon. Scanning was again gated to the ECG waveform during apnea so that any changes in CT number (HU) seen during such an experiment should represent changes due to redistribution of the radiopaque xenon gas.

3.5.2 Repeatability of Regional Washout Measurements

We repeated the multi-breath imaging study (with peak airway pressure of 15 cmH2O for the intervening breaths) three consecutive times to test the repeatability of the regional tracer washout/dilution measurements. During these studies we increased the temporal resolution to include 3 baseline points acquired during ECG gated image acquisition during a brief period of apnea at equilibration followed by gathering the regional time-intensity data over 17 breaths.

3.5.3 Matching with Known Pulmonary Physiology

Matching of the measured ventilation distributions to expected pulmonary physiology was evaluated by ascertaining the presence of vertical (ventral-dorsal axis in a supine pig) ventilation gradients and by repeating the muti-time point imaging protocol with 3 different inspiratory flow rates.

 

 


Figure 4: Data showing the linear relationship between radiopacity and xenon conentration as measured in a plexiglass super-syringe. Data was collected using xenon concentrations of 100%, 80%, 60%,30%, 15% and 0% (room air). Scanner mA and kV settings were 623 and 130 respectively.


Results


4.1 Non-Radioactive Xenon as a CT Contrast Agent

As shown in figure 4, CT x-ray attenuation is linearly related to xenon concentration over the entire range of xenon concentration. Linear regression analysis shows a slope of 1.55 HU enhancement per percentage increase in xenon concentration when calibrated using the EBCT scanner (kV=130, mA=623).

4.2 Single Breath Inhalation

A representative subtraction image obtained from a single breath protocol using 100% xenon as the ventilation tracer gas is depicted in figure 5. In this imaging study, the lung was held at 25 cmH2O airway pressure (i.e. approximately at total lung capacity) during both the xenon-enhanced and un-enhanced scans. The un-enhanced scan was digitally subtracted from the xenon enhanced scan. The subtraction image shows minimal edge effects (some edge effects are seen around the diaphragmatic border) indicating that the enhanced and unenhanced scans were correctly registered. The resulting subtraction image, therefore, should depict the distribution of change in Hounsfield units through the lung field due to a single deep breath of 100% xenon gas and therefore regional ventilation. The mean enhancement from the single deep breath of 100% xenon gas is 321.85 (SE) HU in the lung parenchyma (maximum 50 HU).

  
Figure 5: Subtraction image from an anesthetized pig pre- and post-inhalation of a single breath of 100% xenon gas (both scans at 25 cmH2O airway pressure). There are two major features of this image: 1) The lack of edge misalignment artifact demonstrates our ability to exquisitely control the return to a fixed lung volume following intervening breaths, and 2) The significant enhancement of the lung field after a single breath of 100% xenon gas suggests that a single breath of a high xenon concentration gas mixture provides an adequate signal for regional ventilation measurements to be made.

The results from the temporal stability of the single breath distribution study are depicted in figure 6 with the time-intensity data from representative samples from the dependent (dorsal) and non-dependent (ventral) lung regions in the adjacent panel. Notice that, in the dependent lung areas, the lung density progressively increases over time while the density of areas in the non-dependent lung regions remain relatively stable over the same time. Some of the noise in the time-intensity curves can be attributed to improper cardiac gating by the scanner. At first look, the increase in CT number in the dependent lung regions could indicate that the dense gas xenon is migrating to the dependent (lower regions of the image) regions as time progresses. However, there are no concomitant change in density in the non-dependent regions to indicate such an event. Another possibility is that during the breath hold, oxygen is being taken up faster in the dorsal lung regions (due to a greater blood flow) without an equal exchange for carbon dioxide and thus concentrating the xenon present within the region. Playing the multi-timepoint image set as a cine loop showed that lung volume was diminishing in the dependent regions. We feel that the increase in density over time in the dependent lung regions during a breath hold after inhaling a single breath of 100% xenon is most likely due to the well recognized tendency towards dependent atelectasis during breath hold due to normal oxygen/carbon dioxide exchange characteristics and due to associated concentration of xenon gas.

 

 


Figure 6: Following the indicated regions of interest (ROI's) on the 3mm thin- slice image during apnea reveals appreciable Hounsfield Unit (i.e. density) changes over time. In the dependent regions, the observed changes indicate a monotonically inceasing change in HU over time, whereas the nondependent zones show minor fluctuations in HU over the same time.

4.3 Multi-Breath Washout Protocol

4.3.1 Preliminary Validation of Washout Data

As an initial validation of the equilibrium and washout data obtained from the multi-breath imaging protocol, we used the equation below to predict the equilibrium HU value of each sampled lung region based on regional lung air content, xenon gas concentration used during a given equilibration study and the linear relationship between xenon concentration and HU shown in figure 4.

HUeq = (Airfraction * C * %Xenon) + HUbaseline1

where HUeq is the predicted HU at equilibration, Air fraction is the fraction of the sample that is air[9] (proportional to HUbaseline), C is the slope of the HU vs. [Xe] relationship, and HUbaseline is the HU after all tracer has been cleared from the sample.

This equation was applied to data from washouts studies after equilibration with 33% or 66% xenon gas mixtures. Using measured values of Air fraction=0.52 of a given sample and C=1.55 HU/%Xe, we computed and matched the following predictions to measured observations:

The measured washout data match closely with the predicted values.

4.4 High-Resolution Maps

Figure 7 illustrates the high resolution and detailed regional information obtainable with careful attention to cardiac and respiratory gating during a multi-breath washout protocol. For this imaging study, the lung was equilibrated with 66% xenon gas as the ventilation tracer and the lung was imaged at 0 cmH2O airway pressure (i.e. end expiration or functional residual capacity) at all time points. The intervening breath between consecutive images was to 25 cmH2O airway pressure (i.e. TLC breaths). The first time point acquired is the equilibration image before any tracer has been washed out (diluted) due to inspirations.

 

 


Figure 7: ECG and respiratory gated multi-breath imaging allows sampling washout curves with high-resolution (approximately 6mm by 6mm samples shown) in both dependent and non-dependent lung regions. Clearance curves from the samples depicted indcate that washout occured faster in the dependent lung regions than in the non-dependent lung regions. Note the nearly linear clearance curve for sample 2 in the nondependent lung zone.

Sample regions used to generate data shown in figure 7 represent 6x6 pixel boxes on a 7mm thick slice. In figure 7, change in HU is plotted against breath number during the imaging study. Notice that the sample from the dependent lung region has a faster clearance/dilution of xenon relative to the two samples from the non-dependent region of the lung, suggesting greater ventilation in the dependent regions. An important finding from the time-intensity (concentration) data is that samples in the non-dependent areas of the lung are not cleared (or tracer not fully diluted) of tracer within the ten breaths scanned (see figure 9). Another important point from this image data is that some samples in the non-dependent lung regions have almost linear rather than exponential clearance (dilution) curves as assumed by the Kety model and assumed in all previous x-ray CT assessments of pulmonary ventilation.

We repeated an identical imaging study as above (with peak airway pressure of 15 cmH2O) three consecutive times to test the repeatability of the regional measurements. The data are displayed in figure 8 for four distinct samples in the lung field. To better understand the time it takes lung regions to fully washout to baseline, and to assure a good measure of the equilibrium enhancement, during these studies we increased the temporal resolution to include 3 baseline points acquired (ECG gated) during a brief period of apnea at equilibration followed by gathering the regional time-intensity data over 17 breaths. In all four sample locations depicted in figure 8, xenon dilution curves from the three repeat scans are nearly coincident, indicating that respiratory and cardiac gated image acquisition during a washout study allows reproducible detailed regional ventilation measurements throughout the lungs.

  


Figure 8: Repeatable washout curves were obtained from both dependent and non-dependent lung regions when the same multi-breath imaging protocol was repeated 3 consecutive times. Graphs A-D correspond to ROI's A-D in the image. For each ROI, the 3 washout curves from the thrice repeated experiment have good correspondence with one another, demonstrating that our multi-breath imaging protocol can give repeatable data.

An interesting finding from these repeatability scans is the increase in density seen in the dependent lung zones during the brief apnea at equilibrium (panels B and C of figure 8) that parallels the findings from the single breath study (figure 6). It appears that the lung volume diminishes appreciably within 5-10 seconds of breath hold. Also, note that in panel D, xenon clearance appears linear rather than exponential.

4.5 Vertical Ventilation Gradients

We analyzed regional xenon dilution curves from the dependent and non-dependent lung zones to see whether the relationship between the washouts in the two regions matched with known physiology. Data from 11 regional clearance curves in the dependent lung zone were averaged and plotted on the same scale as the average xenon clearance curve from 11 samples in the non-dependent lung region. As can be seen in figure 9, in the dependent lung region, xenon tracer gas is diluted much more rapidly than the non-dependent lung zones. There is, therefore, evidence of a vertical gradient in ventilation in this supine pig (ventral-dorsal gradient) which is consistent with known pulmonary physiology.

  


Figure 9: To examine the existence of vertical ventilation gradients, we analyzed the relationship between the average clearance curves of 10-11 ROI's placed in random locations (areas containing major airways and blood vessels were avoided) in each the dependent and non-depenedent lung regions in a supine pig. Examining these ROI's during a washout study revealed a significantly steeper washout curve in the dependent lung regions than in the nondependent zones. Notice that the average clearance curve from the nondependent lung region does not appear to have returned to baseline (i.e fully cleared of tracer) within 10 breaths.

4.6 Effect of Inspiratory Flow Rate on Regional Washouts

As seen in figure 10, inspiratory flow rate variations used in this study had no effect on the distribution of air flow (as indicated by xenon washout rates) to non-dependent lung regions in the supine pig. In the dependent regions, however, washout is enhanced at 25 liters/minute inspiratory flow rate while there was no difference between 15 and 20 liters/minute inspiratory flow rate (tidal volume held constant). Since at high flow rates, inertial factors and hence pathway geometry becomes more important in the distribution of air flow (ventilation), our data indicate that the dorsal, diaphragmatic region of the pig lung (dependent lung region in a supine pig) receives an enhanced air flow distribution at higher flow rates perhaps do to favorable airway structure.

 

 


Figure 10: Examining the effects of inspiratory flow rate on ventilation distribution reveals that washout in non-dependent lung regions in the supine pig are not affected by the rate of inspiration over the 15-25 liter/minute inspiratru flow rate range. Tidal volume was held constant in all 3 studies during this experiment. The dependent lung zones of the supine pig, however, show a faster clearance at 25 liters/minute inspiratory flow rate and negligible differences between clearance rates at 15 and 20 liters/minute inspiratory flow.


5 Discussion


Stable (non-radioactive) xenon gas is chemically inert and thus has minimal physiologic effects while providing good radiopacity and contrast as a ventilation tracer during CT imaging. We have implemented a dynamic washout imaging protocol where we image regional tracer gas clearance (time-intensity data) during tidal breathing and a single breath protocol where we subtract a xenon enhanced scan from an identical scan prior to xenon inhalation. The multi-breath washout study provides information on the regional distribution of pulmonary ventilation under dynamic/tidal breathing conditions while the single breath protocol provides information on the distribution of ventilation under static conditions.

 

 


Figure 11: Picture of a pig airway cast. The main airway in a supine pig traverses towards the dorsal (dependent) regions in the base of the lung. We believe the geometric structure of the airways plays an important role in preferential distribution of ventilation to dorsal regions of the pig lung at higher inspiratory flow rates. (From Amirav et. al., J. Appl. Physiol 75(5): 2239-2250, 1993). With permission from the American Physiological Society.

While the single breath protocol is attractive due to the simplicity of the imaging protocol and quantitative image analysis, our results indicate that even with close attention to respiratory and ECG gated image acquisition, density changes begin soon after the onset of a breath hold, confounding assessment of regional ventilation. However, adequate concentrations of xenon (HU changes) are achieved with a single breath approach. Assessment of regional ventilation using a single breath approach, therefore, requires near synchronous volume scanning following a single breath of xenon, currently achievable using the EBCT 50 msec mode of scanning.

While the contaminating density changes during breath hold may not always be as dramatic as shown in figure 6 , they are manifested within the first 5-10 seconds of apnea after a single inhalation of high concentration xenon gas. Because of the time course of this effect, using commercial scanners (other than the EBCT scanner) currently available, where a volumetric image is gathered via table advancements past a fixed scanning plane, use of the single breath ventilation protocol should be limited to volumetric acquisitions taking no longer than 5 seconds. If a volumetric data set of the lungs cannot be obtained synchronously, calculated density changes from digitally subtracting the enhanced and un-enhanced scans of the single breath protocol need to be interpreted with caution. It is of interest to note that this is not a problem using radioactive tracers scans as in nuclear medicine since quantitation in that arena is based on regional radioactive count rates rather than regional density changes.

Unlike previous investigators who have used stable xenon gas in conjunction with x-ray CT imaging for measurement of regional ventilation, we sampled the entire regional tracer clearance curve. Moreover, the previous investigators[14,1,15,16,20,8,6,5] were unable to use cardiac gated image acquistion because of long scan times. One study did use an EBCT scanner, however, they did not report cardiac gating[14]. Moreover, the previous studies imaged only a very limited number of time points and assumed the clearance or wash-in curves to be mono-exponential.

Accurate sampling of the entire washout curve indicates that more than 10 breaths are needed to clear tracer from regions with low ventilation. Of even greater importance, is the finding from our studies that some non-dependent lung regions demonstrate nearly linear rather than mono-exponential clearance curves as assumed in all previous investigations of regional pulmonary ventilation using x-ray CT and stable xenon gas. Measurements based on limited time points modeled as a mono-exponetial may not, therefore, yield physiologically accurate information.

At high flow rates, inertial factors and hence pathway geometry become more important in the distribution of air flow (ventilation). Our data indicate that the dorsal, diaphragmatic region of the pig lung (dependent lung region in a supine pig) receives an enhanced air flow distribution at higher flow rates, perhaps do to favorable airway structure. A pig airways cast, shown in figure 11, illustrates that due to the monopodial branching pattern of the pig airways, there is a major airway traversing to the dorsal, diaphragmatic region of the pig lung. It is our hypothesis that this airway branching pattern when coupled with high flow rates would be expected to produce greater tunrover rates in the dependent lung region of a supine pig. Functional x-ray CT imaging may, therfore, allow us to explore pulmonary structure-function relationships in vivo.


6 Conclusions


Our preliminary results indicate that, with careful attention to cardiac and ventilatory gating during image acquisition, x-ray CT based imaging of stable xenon gas can yield detailed maps of regional lung function. Single breath methods are possible if scanning times are rapid. Additional work is needed to evaluate the prevelance of near linear washout regions before we are able to recommend appropriate sampling intervals and total sampling duration. Used in conjunction with thin-slice volumetric imaging contrast injection for perfusion imaging[10], x-ray CT can provide here-to-fore unavailable information relating pulmonary structure and function.

We are currently automating the image analysis and integrating ventilation measurements into our customized image analysis package[10] to allow rapid analysis of the clearance curves and construction of a corresponding accurate and quantitative color coded functional maps. Our future investigations include further validation of the washout data and developing a model-free method of calculating regional ventilation based on the measured breath to breath tracer dilution for a regional sample and is thus free from possible oversimplfying assumptions.


Acknowledgements: We would like to thank Steve DeJong and Tony Smith for their help with animal preparation. This study was funded in part by a grant from Mallinckrodt Medical, Inc. and NIH-HL-42672. Xenon gas was generously donated by PraxAir.


7 References


1
Awai, K., K. Ito, S. Katsuta, K. Fujikawa, S. Nakamura, T. Ichiki, K. Yamane, M. Fujie, K. Hasegawa, O. Nishida, et al. ``Xenon enhanced dynamic computed tomography of the lung; regional ventilation measurement''. Rinsho Hoshasen--Japanese Journal of Radiology, 35(7):847--853, 1990.

2
Ball, W. C., P. B. Stewart, L. G. S. Newsham, and D. V. Bates. ``Regional pulmonary function studied with xenon133''. J Clin Invest, 41(3):519--631, 1962.

3
Boyd, D. P. and M. J. Lipton. ``Cardiac computed tomography''. Proceedings of the IEEE, 71:298--307, 1983.

4
Cullen, S. C. and E. G. Gross. ``Anesthetic properties of xenon in animals and human beings with additional observation on krypton''. Science, 113:580--581, 1951.

5
Gur, D., B. P. Drayer, H. S. Borovetz, B. P. Griffith, R. L. Hardesty, and S. K. Wolfson. ``Dynamic computed tomography of the lung: regional ventilation measurements''. J Comput Assist Tomogr, 3(6):749--753, 1979.

6
Gur, D., L. Shabason, H. S. Borovetz, D. L. Herbert, G. J. Reece, W. H. Kennedy, and C. Serago. ``Regional pulmonary ventilation measurements by xenon enhanced computed tomography: an update''. J Comput Assist Tomogr, 5(5):678--683, 1981.

7
Gur, D., H. Yonas, and W. F. Good. ``Local cerebral blood flow by xenon-enhanced CT: current status, potential improvements, and future directions''. Cerebrovasc Brain Metab Rev, 1(1):68--86, 1989.

8
Herbert, D. L., D. Gur, L. Shabason, W. F. Good, J. E. Rinaldo, J. V. Snyder, H. S. Borovitz, and M. C. Mancini. ``Mapping of human local pulmonary ventilation by xenon enhanced computed tomography''. J Comput Assist Tomogr, 6(6):1088--1093, 1982.

9
Hoffman, E. A. ``Effect of body orientation on regional lung expansion: a computed tomographic approach''. J Appl Physiol, 59:468--480, 1985.

10
Hoffman, E. A., J. K. Tajik, and S. D. Kugelmass. ``Matching pulmonary structure and perfusion via combined dynamic multislice CT and thin-slice high-resolution CT''. Computerized Medical Imaging and Graphics, 19(1):101--112, 1995.

11
Holden, J. E., S. S. Winkler, D. C. Flemming, W. R. Ip, and J. F. Sackett. ``Imaging methods in the transmission computed tomographic measurement of regional xenon kinetics in the brain''. Radiology, 135:501--505, 1980.

12
Kety, S. S. ``The theory and applications of the exchange of inert gas at the lungs and tissues''. Pharmacol Rev, 3:1--42, 1955.

13
Kishore, P. R., G. U. Rao, R. E. Fernandez, R. L. Keenan, G. D. Arora, P. Gadisseux, L. M. Stewart, A. O. Wist, P. P. Fatouros, D. Dillard, et al. ``Regional cerebral blood flow measurements using stable xenon enhanced computed tomography: a theoretical and experimental evaluation''. J Comput Assist Tomogr, 8(4):619--630, 1984.

14
Murphy, D. M., J. T. Nicewicz, S. M. Zabbatino, and R. A. Moore. ``Local pulmonary ventilation using nonradioactive xenon-enhanced ultrafast computed tomography''. Chest, 96(4):799--804, 1989.

15
Snyder, J. V., B. Pennock, D. Herbert, J. E. Rinaldo, J. Culpepper, W. F. Good, and D. Gur. ``Local lung ventilation in critically ill patients using nonradioactive xenon-enhanced transmission computed tomography''. Crit Care Med, 12(1):46--51, 1984.

16
Tomiyama, N., N. Takeuchi, H. Imanaka, N. Matsuura, S. Morimoto, J. Ikezoe, T. Johkoh, J. Arisawa, and T. Kozuka. ``Mechanism of gravity-dependent atelectasis. analysis by nonradioactive xenon-enhanced dynamic computed tomography''. Invest Radiol, 28(7):633--638, 1993.

17
Winkler, S. S., J. E. Holden, J. F. Sackett, D. C. Flemming, and S. C. Alexander. ``Xenon and krypton as radiographic inhalation contrast media with computerized tomography: Preliminary note''. Invest Radiol, 12(1):19--20, 1977.

18
Winkler, S. S., J. F. Sackett, J. E. Holden, D. C. Flemming, S. C. Alexander, M. Madsen, and R. I. Kimmel. ``Xenon inhalation as an adjunct to computerized tomography of the brain: Preliminary study''. Invest Radiol, 12(1):15--18, 1977.

19
Wist, P. P. F. A. O., P. R. Kishore, D. S. DeWitt, J. A. Hall, R. L. Keenan, L. M. Stewart, A. Marmarou, S. C. Choi, and H. A. Kontos. ``Xenon/computed tomography cerebral blood flow measurements. methods and accuracy''. Invest Radiol, 22(9):705--712, 1987.

20
Yonas, H., W. F. Good, D. Gur, S. K. W. Jr, R. E. Latchaw, B. C. Good, R. Leanza, and S. L. Miller. ``Mapping cerebral blood flow by xenon-enhanced computed tomography: clinical experience''. Radiology, 152(2):435--442, 1984.





©1994-2000 Division of Physiologic Imaging, Dept. of Radiology, Univ. of Iowa


Papers | DPI Homepage | VIDA | NLM | Contact Us | Search



Last modified: Fri May 26 16:10:38 CDT